  1. Check settings Please make sure that the fogging settings have been made correctly. Instructions for correct operation can be found in the following video:
  1. Check internal connection If the settings are correct, but the nebulizer still does not start, it may be that the internal nebulization line has been disconnected.
In this case, please contact our Customer Serviceto obtain technical support.
The normal oxygen content of ambient air is 21%. If the display of your Pet ICU does not show 21%, a Manual calibration required. Please watch our calibration video: 为了 KANISTAR宠物重症监护室 the function for controlling the high oxygen concentration is a central feature. An inaccurate display can interfere with monitoring your animal’s oxygen supply. The calibration shown in the video ensures that the displayed oxygen value is correct and that the ICU is working according to its function.
Please watch our instruction video first to make sure the oxygen settings are correct: If the oxygen level rises only slowly and does not exceed 37%, please note the following points:
  1. Switch from external to internal circulation.
  2. Close the small door on the front cover Cover it with the attached Plastic cover
  3. Close the following openings:
    • Connection for humidifier
    • Connection for nebulizer
    • Urine drainage connection
  4. Turn the flow meter to 10L.
These steps ensure that the ICU works hermetically and an oxygen concentration of over 37% can be achieved.